A handover of a 50–75 GHz WR15 bench should leave the next operator an identifiable RF path, source envelope and control state. The amplifier model alone cannot reproduce the previous run. Keep the setup drawing, calibration basis, cooling arrangement and baseline files together.
The CorelixRF CRF-PA-50000M75000M-8W is a millimeter wave power amplifier configuration whose June 2026 V1.0 electrical table lists 8 W minimum Output Power and 39 dB minimum small-signal gain. It specifies front-panel WR15 input and output, a 19-inch 3U enclosure, forced-air cooling, and RS485 and LAN control. The output row does not define P1dB, Psat or a modulated-waveform linear-power limit.
Identify both WR15 paths
Record the source-side extension, waveguide sections, bends, transitions and attenuation separately from the output-side coupler, load, fixture or antenna path. Identify the amplifier input and output planes and the point where the project requires power. A measurement after a coupler or fixture must retain the applicable path corrections.

Obtain the final flange and handling instructions. Support rigid sections independently, leave fastening and inspection access, and keep mating surfaces protected when disconnected. A WR15 designation does not establish every flange detail or the power and frequency capability of the whole assembly.
Leave an approved source recipe
Save the source frequencies, waveform, input limits, gain commands and startup/inhibit procedure approved for the delivered configuration. The 39 dB small-signal gain does not establish full-power drive or maximum permitted input. The 15 dB maximum gain-control range does not specify the usable adjustment, step size or level accuracy.
The next operator needs the actual source envelope and command behavior, including what happens after a host restart, alarm or connection loss. Do not infer that the amplifier disables the upstream source whenever a protection event occurs.

Make cooling and telemetry part of the record
The source lists AC 220 V ±10%, 50/60 Hz and operation from 0 to +50 °C. Obtain final dimensions, mass, inlet/exhaust directions and required clearances before fixing the 3U rack position. Retain nearby heat sources, rack restrictions and run duration with the setup.
Temperature and current monitoring are listed; forward/reverse power monitoring is optional. Confirm fitted channels, units, update behavior and protocol. The named over-temperature, over-drive, over-voltage and VSWR protections do not supply numerical trip thresholds, response times or recovery conditions. Leave the approved operator response with the station.
Keep a usable first-run baseline
| Handover item | Record to retain |
|---|---|
| Hardware configuration | Full model, unit identity, fitted options and drawing revision |
| RF arrangement | Both WR15 paths, load identity, reference planes and corrections |
| Operating recipe | Frequency, waveform, source envelope, gain state and timing |
| Installation and control | Supply/cooling conditions, monitored fields, alarms and recovery |
| Baseline result | Raw readings, method, conditions, applicable limits and unresolved items |

Distinguish source specifications from measurements made on this setup. The RF amplifier testing guide helps define the result fields; it does not demonstrate that the station has passed them.
Record changes before comparing runs
A replaced transition, moved rack, changed load or revised control recipe should produce a new setup record. Recheck the affected path, calibration, cooling or control behavior before treating a new reading as comparable with the earlier baseline. A change-triggered check does not by itself mean the amplifier is defective.
If the band or interface requirement changes, compare the millimeter-wave configurations before substituting a unit. Send the WR15 handover arrangement for review with the source recipe, output plane, load, cooling conditions, fitted functions and the unresolved behavior.