Choose an RF amplifier for a mismatched load by defining the required behavior at the output port: continue producing the specified signal, reduce power, or shut down and recover under agreed conditions. Then compare documented amplifier configurations against that requirement. Input VSWR, saturated power and a protection label cannot substitute for this output-load decision.

Separate input matching from output tolerance
Input VSWR describes the match presented at the amplifier input under its stated conditions. It does not specify how the output stage survives reflected power. An output mismatch claim needs its own test conditions; do not turn an input-match value into a load-survival rating.
Define the output reference plane and the cable, coupler, filter or other components between it and the load. Forward and reverse readings taken at another plane cannot be silently treated as amplifier-port readings. Record the measurement method and its applicable calibration, directivity and dynamic-range limits.

Reflected power calculation: estimate waves, not survival
For a single-frequency steady-state wave in the same real reference impedance, the reflection-coefficient magnitude is |Γ| = (VSWR − 1)/(VSWR + 1). The reflected-to-forward power ratio is |Γ|². These relations describe the waves at a defined plane; they do not specify the amplifier’s safe operating boundary.
As a calculation only, VSWR 2:1 gives |Γ| = 1/3 and a reflected-power fraction of about 11.1%. If the actual forward power at that same plane is 100 W, the corresponding reflected power is about 11.1 W. This is not a measurement of a CorelixRF unit or a claim that a particular output stage can tolerate that condition. For the related system or selection decision, see RF Amplifier Selection: Decide What the Evidence Supports.
A protected amplifier may reduce forward power as the mismatch develops. Use the forward power measured during the condition rather than continuing to calculate from its nameplate output. Phase and the amplifier’s output network also matter to the internal stress, so the reflected-power fraction alone is insufficient for a survival decision.
Define the output mismatch requirements before comparing models
Provide the required frequency range, waveform, output-power definition and thermal conditions. For pulsed operation, include the pulse conditions relevant to the application. A saturated-power label does not establish linear modulated output, and a brief condition does not establish continuous operation.
Ask how the permitted mismatch is defined: its magnitude, relevant phase coverage, duration, applied power, frequency and cooling condition. State whether the requirement is continued output, protective reduction, shutdown without damage or a documented recovery sequence. These are different acceptance outcomes.

| Requirement | Information the supplier must define |
|---|---|
| Output must remain usable | Signal-quality requirement and permitted output change during mismatch |
| Power reduction is acceptable | Foldback condition, resulting output and any thermal constraints |
| Shutdown is acceptable | Trip condition, fault indication and whether restart is automatic or commanded |
| Qualification evidence is requested | Sample/configuration identity, mismatch magnitude and phase, frequency, duration, power, temperature and acceptance outcome |
Read protection descriptions as functions to verify
A VSWR alarm or protection listing does not define a universal threshold, response time or recovery policy. Request the behavior for the actual delivered configuration, including what is monitored and what happens when the condition clears. A telemetry update interval is not automatically the protection response time. Use the RF power amplifier platforms page to compare the product family, then request the exact configuration and evidence required by this review.
Keep any mismatch demonstration within the supplier-approved procedure and instrument, load and interconnect limits. Do not improvise an open- or short-circuit test from a general article. Retain the sample identity, settings, traces, acceptance criteria and any damage or recovery observations with the result.
Mechanical and environmental evidence remains a separate selection input. An enclosure photograph or a successful vibration report does not establish RF mismatch tolerance. Likewise, a mismatch test does not qualify the installation for its mechanical environment. Match each requirement to the configuration and evidence that actually addresses it.
Prepare a comparison sheet with the requirement, the proposed configuration, the supporting document and the remaining evidence gap for each option. Use the RF testing documentation route to agree the required records.
Send CorelixRF the output mismatch requirement. Include waveform, output-power definition, frequency, load conditions and the required protection outcome.