A BCI calibration curve shows 20 W. Does that mean a 20 W RF amplifier is enough? First check where that power is defined, whether path loss is already included and how the amplifier will operate. Those answers turn a curve point into a useful requirement—not yet a model recommendation.
For EMC engineers and technical buyers, this guide turns bulk current injection calibration data into a short request for quotation (RFQ). It starts from an existing record and the method/version in your test plan; it is not a general conducted-immunity test procedure.
1. Before comparing quotes, explain what the curve means
A column marked only “drive power” is not enough for a quotation. Does it mean the signal-generator setting, the amplifier output, or forward power at the injection-probe input? These are different quantities. Identify where the power is defined—its reference plane—before comparing amplifier quotes.
Ask whoever produced the record to identify the quantity, its units and its reference plane. Keep the probe and calibration-fixture identities with the method and version used for that record.
| Recorded quantity | Question to resolve before using it |
|---|---|
| Signal-generator setting | What does that setting represent, and where is the corresponding RF power defined? |
| Amplifier output power | Is this already at the amplifier output reference plane? Which corrections have been applied? |
| Forward power at the probe input | What path lies between that plane and the amplifier output, and has its loss already been compensated? |
Keep the calibration-fixture record separate from the actual harness test. The fixture result alone does not guarantee harness current; retain the configuration and control method for each.

2. What does a 20 W curve point mean for the amplifier?
Illustrative Example — not a product test or compliance result.
For this example only, assume that:
- the curve records forward power at the injection-probe input reference plane;
- the path from the amplifier output to that plane has a known, constant 1.0 dB loss;
- the path is linear and the loss was established under the same matched conditions;
- that loss has not already been compensated in the curve.
Under these assumptions, the converted power inputs are shown below. The formula is available if you need to check the calculation:
Optional: see the conversion formula
Both powers are in watts; L is the path loss in dB. The factor for 1.0 dB is approximately 1.2589.
| Illustrative record point | Probe-input forward power | Path loss | Converted amplifier-output-plane power |
|---|---|---|---|
| A | 8 W | 1.0 dB | 10.07 W |
| B | 20 W | 1.0 dB | 25.18 W |
| C | 12 W | 1.0 dB | 15.11 W |
A, B and C are illustrative record points, not measured frequencies. Apply the stated path-loss correction once; if it is already included in the curve, adding it again would double-count it. Use the calculation only when its assumptions hold.
What happens after the calculation? Put 25.18 W into the requirements sheet as a conditional amplifier-output input. It is not evidence that a 30 W amplifier will pass. Next, review the proposed configuration under the required waveform, load, duration, cooling and protection conditions using the steps below.

3. Turn the curve into a configuration review
- Identify the record. Keep the quantity, reference plane, frequencies and probe/fixture identities together.
- Carry over the conditional power input. Apply only the path correction supported by the record, keeping its assumptions attached.
- Review a candidate against the operating conditions. Ask for configuration-specific evidence at the required frequencies and waveform, rather than comparing a nominal power label with the curve peak.
- Resolve the gaps before selecting a model. Record the open conditions and which party will confirm them.
Use this compact requirements table for that discussion:
| Requirement to attach | Question for the configuration review |
|---|---|
| Operating band and curve points | How will the proposed configuration be assessed across the required frequencies? |
| Power quantity, reference plane and corrections | Which values are amplifier-output inputs, and which corrections are already included? |
| Waveform and operating duration | What evidence is needed for the requested output under these conditions? |
| Load, reflection and protection limits | Which conditions and restrictions must the proposed configuration address? |
| Cooling and supply responsibilities | What installation information does each party need to confirm? |
| Control and integration requirements | Which operating and interface conditions remain open? |
State whether the plan uses substitution or a power-limited closed-loop method. Keep their control and limit requirements separate; another procedure’s multiplier or stopping threshold is not a substitute for your plan.
For detailed path and installation questions, use the EMC RF-path planning guide and the EMC amplifier supply and cooling checklist. Their model-specific ratings remain tied to their own configurations.

4. Start with three things you already have
You do not need a completed report for every open condition before starting a discussion. Send the records you already have, identify the gaps and ask which ones must be resolved before a configuration can be recommended.
- The existing curve: include its recorded quantity, units, reference plane and any corrections already applied.
- The test identity: give the operating band, method/version and available probe and fixture information.
- The known operating conditions: provide the waveform, run duration and any load, cooling or control constraints. Mark unknown items as awaiting confirmation.
Use the CorelixRF EMC amplifier range as the product starting point for a hardware requirements review. If the request includes the wider RF path, the EMC and laboratory RF system scope helps frame the integration discussion. Keep the discussion tied to the recorded conditions and the evidence needed for the proposed configuration.
Optional: expand the request before comparing quotations
- Keep the source record, method/version and probe/fixture identities together.
- List the quantity, reference plane and correction status for each power column.
- Separate known path loss from losses or load conditions still requiring review.
- State the waveform, duration and applicable power/reflection limits.
- Assign control, supply and cooling questions to the relevant parties.
- Ask for configuration-specific evidence against the agreed requirements; leave unsupported claims open.
Need an RF amplifier for your BCI setup? Request a BCI amplifier requirements review with CorelixRF. Share your operating band, calibration record and known conditions. Use the discussion to clarify the power input, identify open requirements and agree what configuration-specific evidence is needed.
The scope is RF hardware assessment, not certification or third-party testing. A calibration curve alone does not establish test compliance.